Setting up the PCR Reaction - NEBNext Multiplex Oligos for Illumina (Dual Index Primers Set 2) (E7780)

Protocol

1.1. PCR Amplification

For < 96 samples, follow the protocol in Section 1.1A. For 96 samples, follow the protocol in Section 1.1B.

1.1A. Setting up the PCR reactions (< 96 samples)

Note: We recommend using a PCR work-up rack such as the TruSeq Index Plate Fixture (Illumina #FC-130-1005) to assist in properly combining the index primers during the PCR amplification step. Alternatively, 96-well deep well plates can be used and aligned against a PCR plate as in the diagram below.

1.1A.1. Ensure that a valid combination of i7 and i5 primers is used. See Appendix A to verify that correct primer combinations have been selected.

1.1A.2. Arrange the index primers in the Index Plate Fixture as follows:

  1. Arrange the (green) i7 primers in increasing order horizontally, so that the lowest number i7 index primer is in column 1, second lowest number i7 index primer is in column 2, etc. 
  2. Arrange the (white) i5 primers in increasing order vertically, so that the lowest number i5 index primer is in row A, second lowest number i5 index primer is in row B, etc.
  3. Record their positions on the PCR setup template (see Appendix B).

1.1A.3. Using a multichannel pipette, add desired volume of (white) i5 primers to every column (as needed) of the PCR plate. It is critical to change tips between columns to avoid cross-contamination.

1.1A.4. Discard the original i5 white caps and apply new caps to avoid index cross-contamination

1.1A.5. Using a multichannel pipette, add desired volume of (green) i7 primers to every row (as needed) of the PCR plate. It is critical to change tips between rows to avoid cross-contamination.

1.1A.6. Discard the original i7 orange caps and apply new caps to avoid index cross-contamination.

1.1A.7. Add 25 μl (blue) NEBNext PCR Master Mix to each well that contains primers.

1.1A.8. Add desired volume of adaptor ligated DNA for a final volume of 50 µ l to the corresponding well. Gently pipette up and down 5–10 times to mix. It is critical to change tips between samples to avoid cross-contamination. Record each sample position on the PCR setup template.

1.1A.9. Cover the plate with Bio-Rad Microseal "A" Film, and seal with a rubber roller. Quickly centrifuge.

1.1A.10. Perform PCR according to recommended cycling conditions.

 

1.1B. Setting up the PCR reactions (96 samples)

Note: We recommend using a PCR work-up rack such as the TruSeq Index Plate Fixture (Illumina #FC-130-1005) to assist in properly combining the index primers during the PCR amplification step. Alternatively, 96-well deep well plates can be used and aligned against a PCR plate as in the diagram below.


1.1B.1. Arrange the index primers in the Index Plate Fixture as follows:

a. Arrange  (green) i7 primers in increasing order horizontally, so that i713 is in column 1, i714 is in column 2, i715 is in column 3, etc.
b. Arrange the  (white) i5 primers in increasing order vertically, so that i509 is in row A, i510 is in row B, i511 is in row C, etc.
c. Record their positions on the PCR setup template (see Appendix B).

1.1B.2. Using a multichannel pipette, add desired volume of  (white) i5 primers to every column of the PCR plate. It is critical to change tips between columns to avoid cross-contamination.

1.1B.3. Discard the original i5 white caps and apply new caps to avoid index cross-contamination.

1.1B.4. Using a multichannel pipette, add desired volume of  (green) i7 primers to every row of the PCR plate. It is critical to change tips between rows to avoid cross-contamination.

1.1B.5. Discard the original i7 orange caps and apply new caps to avoid index cross-contamination.

1.1B.6. Add 25 μl  (blue) NEBNext PCR Master Mix to each well.

1.1B.7. Add desired volume of adaptor ligated DNA for a final volume of 50 µl to the corresponding well. Gently pipette up and down 5–10 times to mix. It is critical to change tips between samples to avoid cross-contamination. Record each sample position on the PCR setup template (see Appendix B).

1.1B.8. Cover the plate with Bio-Rad Microseal "A" Film, and seal with a
rubber roller. Quickly centrifuge.

1.1B.9. Perform PCR according to recommended cycling conditions. 

 

Appendix A: Principle for Use and Pooling Guide

The Principle of Dual Index Primers

The dual index primer strategy utilizes two 8 base indices within each primer; i7 primers contain indices that are adjacent to the P7 sequence; i5 primers contain indices that are adjacent to the P5 sequence. Dual indexing is enabled by adding a unique index to both ends of a sample to be sequenced. Up to 96 different samples can be uniquely indexed by combining each of the 12 i7 primers with each of the 8 i5 primers. Similarly, < 96 samples can be uniquely indexed by combining i7 primers with i5 primers as follows:

N = Number of samples = X(i7) * Y(i5) + other primers as needed

Examples:

1. For N = 12 samples

Option 1:

4 (i7) * 3 (i5) From the i7 primers, choose a valid set of 4. From the i5 primers choose a valid set of 3. Use each i7 primer with each i5 primer to form 12 primer pairs for PCR amplification of 12 libraries. When setting up the sequencing run, select “Dual Index” and choose the indices used for each sample.

Option 2:

3 (i7) * 4 (i5) From the i7 primers, choose a valid set of 3. From the i5 primers choose a valid set of 4. Use each i7 primer with each i5 primer to form 12 primer pairs for PCR amplification of 12 libraries. When setting up the sequencing run, select “Dual Index” and choose the indices used for each sample.

Option 3:

6 (i7) * 2 (i5) From the i7 primers, choose a valid set of 4 and add any other two i7 primers, for a total of 6 primers. From the i5 primers choose a valid set of 2. Use each i7 primer with each i5 primer to form 12 primer pairs for PCR amplification of 12 libraries. When setting up the sequencing run, select “Dual Index” and choose the indices used for each sample. Option 4: 12 (i7) * 1 (i5) Use all 12 i7 primers. Use any i5 primer. Use each i7 primer with the i5 primer to form 12 primer pairs for PCR amplification of 12 libraries. When setting up the sequencing run, select “Single Index”, and choose the i7 index used for each sample. CHAPTER 2 12

2. For N = 26 samples

Option 1:

6 (i7) * 4 (i5) + 2 (i5) From the i7 primers, choose a valid set of 4 and add any other two i7 primers, for a total of 6 primers. From the i5 primers choose a valid set of 4 and add any other two i5 primers, for a total of 6 primers. Use each i7 primer with four of the i5 primers to form 24 primer pairs. Use any of the six i7 primers with the remaining two i5 primers to form 2 primer pairs. This will give you a total of 26 primer pairs for PCR amplification of 26 libraries. When setting up the sequencing run, select “Dual Index” and choose the indices used for each sample.

Option 2:

6 (i7) * 5 (i5) From the i7 primers, choose a valid set of 4 and add any other two i7 primers, for a total of 6 primers. From the i5 primers choose a valid set of 4 and add any other one i5 primer. Use each i7 primer with each i5 primer to form 30 primer pairs for PCR amplification. Use 26 of the 30 primer pairs to amplify 26 libraries. When setting up the sequencing run, select “Dual Index” and choose the indices used for each sample.

Low Plexity Pooling Guidelines

Illumina uses a red laser/LED to sequence A/C and a green laser/LED to sequence G/T. For each cycle, both the red and the green channel need to be read to ensure proper image registration (i.e. A or C must be in each cycle, and G or T must be in each cycle). If this color balance is not maintained, sequencing the index read could fail. The following table lists some valid combinations that can be sequenced together. Note: for 1-plex (no pooling), use any i7 primer with any i5 primer. In this case it is important to select “0” index reads in the Illumina Experiment Manager.

CAUTION: Sufficient primers are provided to generate 96 different samples if each i5 primer is used only once with each i7 primer.

If using subsets of i5 and i7 primers multiple times, you may have to readjust primer pairs to be able to generate 96 samples.

 

Table 2.1. Pooling: 2–12 libraries; Sequencing Workflow: Single Index (Select "1" Index Reads in the Illumina Experiment Manager)

PLEX i7 PRIMERS i5 PRIMERS
2 i713 and i721
i714 and i724
i715 and i716
i717 and i719
i718 and i720
i722 and i723
Any i5 Primer
3 i714, i715 and i716
i715, i716 and i717
i717, i718 and i719
i718, i719 and i720
i721, i722 and i723
i722, i723 and i724
Any i5 Primer
4 i713, i714, i715 and i716
i715, i716, i717 and i718
i717, i718, i719 and i720
i718, i719, i720 and i721
i720, i721, i722 and i723
i721, i722, i723 and i724
Any i5 Primer
5 -12 Any valid i7 4-plex with any other i7 Primers Any i5 Primer

 

Table 2.2. Pooling: 6+ libraries; Sequencing Workflow: Dual Index (Select "2" Index Reads in the Illumina Experiment Manager).

PLEX i7 PRIMERS i5 PRIMERS
6-12 Any 2 plex combination from Table 2.1 with and any other i7 primer(s) (as needed) i509, i511 and i513
i510, i511 and i513
i511, i513 and i514
i511, i513 and i515
i511, i515 and i516
Greater than 12 Any 3 plex combination from the Table 2.1 with any other i7 primer(s) (as needed)
i509, i510, i511, i513 and any other i5 primer (as needed)
i509, i511, i512, i513 and any other i5 primer (as needed)
i509, i511, i514, i515 and any other i5 primer (as needed)
i510, i511, i515, i516 and any other i5 primer (as needed)
i511, i513, i515, i516 and any other i5 primer (as needed)

Tables 2.1. and 2.2. do not include a complete list of all valid index combinations. Please check the sequences of each index to be used to ensure that you will have signal in both the red and green channels for every cycle. See example below: